Case study: Stable pH measurement for reliable water treatment

pH measurement in polymer-containing process streams is one of the most challenging applications in industry. At BASF in Heerenveen, a producer of water-based binders for the printing ink and coatings industry, unstable pH readings caused major disruption to the ultrafiltration (UF) plant. The results were clear: insufficient thickening of polymers, reduced process efficiency and a high maintenance burden for the team.
The problem was partly due to the nature of the process stream itself. The polymers formed films and adhered to surfaces, including pH sensors. This quickly fouled the sensors and made measurements unreliable. Without continuous monitoring, pH deviations went unnoticed for long periods, with serious consequences for the UF plant.

BASF was looking for a solution that could provide reliable pH data continuously upstream of the UF plant, allowing operators to make timely adjustments through dosing. The choice was an inline pH system with automatic cleaning. The system comprises a Knick Stratos Multi transmitter, a Knick UniClean 720 and a pneumatically actuated retractable assembly. With this setup, the sensor is automatically retracted from the process line and cleaned without interrupting the process.
The results are convincing. pH measurement remains stable and reproducible, even in a polymer-containing stream. BASF Nederland Asset Manager Mark Lazonder confirms: “Even after six months, recalibration was still not necessary, and the maintenance team now has to intervene manually only rarely.” The UF plant is more reliable, the water treatment process is more stable and the staff’s workload has decreased.
A business enquiry for ELSCOLAB? Get in touch
Companies in this story
Keep following this
Get an email when a new story is published. No account needed.
- ThemeKeeping processes stable and clean
- SubjectNews
- CompanyELSCOLAB